EasyManua.ls Logo

cobas c 111 - Revision 7: Replace Sample Area Fan by Display Module Fan

cobas c 111
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
Addendum version 1.0 · 11
cobas c 111 analyzer Revisions to cobas c 111 instrument publications
Revision 7: Replace "sample area fan" by "display module fan"
Revision 7: Replace "sample area fan" by "display module fan"
System status Check the texts for the status of hardware items and on IDs of installed software.
Publication title Publication version Software version Publication date Revised topic title
Operators Manual 4.1 4.1 October 2013 Software > Key screens > System status
Table 8 Revision location
Status of the instrument.
The main cover is open.
A development channel application is run without using extra wash cycles.
o Roche strongly recommends to always use extra wash cycles with
development channel applications, and also to always load extra cleaner
when tests with extra wash cycles are used.
o Roche Diagnostics Ltd. assumes only limited liability when using the
cobas c 111 instrument in conjunction with the cobas c 111
Development Channel Programming Software. For detailed information
on this matter refer to the latest version of the Development Channel
Registration Form cobas c 111 and the cobas c 111 Development
Channel Operator’s Manual.
Temperature status for reagent cooler and cuvette ring.
The temperature is outside the acceptable range.
Status of the display module fan
The fan is not running.
Display information on the fill status of each of the external bottles.
The color of the underlying buttons is displayed.
See Checking the external bottles on page A-96.
Display the maintenance actions list.
The color of the most urgent maintenance action is displayed.
See Maintenance on page A-110.
Status of the printer paper.
The printer is out of paper.