Roche Diagnostics
Addendum version 1.0 · 17
cobas c 111 analyzer 1
Revisions to cobas c 111 instrument publications
Revision 13: Note for the red Alarm LED interpretation added
Checking for alarm messages To check for problems during processing
1 Observe the Alarm LED.
2 Take appropriate action.
For details on dealing with alarm messages, see Alarm monitor on page D-6.
Publication title Publication version Software version Publication date Revised topic title
Operator’s Manual 4.1 4.1 October 2013 Daily operation > Analyzing samples
> Checking for alarm messages
Table 14 Revision location
LED
Meaning Possible action
No color (Off)
There are no unconfirmed
alarm messages.
No action is required.
Yellow
There is at least one
unconfirmed alarm message.
User intervention is required as soon
as possible. Processing can continue
for the time being.
Check the details of the message.
Red
There is at least one
unconfirmed alarm message.
Immediate user intervention is
required. Processing may not be able
to continue without it.
Check the details of the message.
Note: When the alarm message
results from a timing error, all tests
run or results obtained after the time
of the error time stamp should be
retested.
An acoustic signal is sounded when an alarm is generated. You can adjust the volume
(Utilities > Configuration > System > Volume).