Roche Diagnostics
20 Addendum version 1.0 ·
1 cobas c 111 analyzer
Revisions to cobas c 111 instrument publications
Revision 16: Note for reacting to red alarm messages added
Reacting to alarm messages To react to an alarm message
1 Observe the Alarm LED.
Publication title Publication version Software version Publication date Revised topic title
Operator’s Manual 4.1 4.1 October 2013 Messages and alarms > Alarm monitor
> Reacting to alarm messages
Table 17 Revision location
LED
Meaning
No color (Off)
There are no unconfirmed alarm messages.
Yellow
There is at least one unconfirmed alarm message. You need to deal
with it as soon as possible.
Red
There is at least one unconfirmed alarm message. You need to deal
with it as soon as possible.
Note: When the alarm message results from a timing error, all tests
run or results obtained after the time of the error time stamp should
be retested.
An acoustic signal is sounded when an alarm is generated. You can adjust the volume
(Utilities > Configuration > System > Volume).