cobas c 111 analyzer Revisions to cobas c 111 instrument publications
Revision 3: Note for color interpretation for alarm button added
Revision 3: Note for color interpretation for alarm button added
Interpreting the alarm LED
Publication title Publication version Software version Publication date Revised topic title
Operator’s Manual 4.1 4.1 October 2013 Software > Working with the user
interface > Messages > Alarm monitor
Table 4 Revision location
No color, off
There are no unconfirmed alarm messages.
Yellow
There is at least one unconfirmed alarm message. You need to
deal with it as soon as possible.
Red
There is at least one unconfirmed alarm message. You need to
deal with it immediately, processing may not be able to continue
unless you do so.
Note: When the alarm message results from a timing error, all
tests run or results obtained after the time of the error time
stamp should be retested.
Acoustic signal
An acoustic signal is sounded when an alarm is generated. You
can adjust the volume
(Utilities > Configuration > System > Volume).