Revision 10: Additional information for preparing cuvettes in the short guide
Revision 10: Additional information for preparing cuvettes in the short guide
Preparing the system The following table provides an overview of the steps that make up the preparation
process.
Publication title Publication version Software version Publication date Revised topic title
Operators Manual 4.1 4.1 October 2013 Daily operation
> Preparing the system
> Short guide
Table 11 Revision location
6
Prepare the cuvettes. To avoid timing errors ensure that all available cuvette
segment positions are filled.
Always replace any completely used cuvette segments. A
partially used cuvette segment does not have to be replaced.
Replace all red cuvette segments.
1. Press the cuvette button.
2. Open the main cover.
3. Replace the cuvette segment.
4. Press to confirm the replacement.
5. Replace the remaining segments that need replacing.
6. Press to proceed to the next stage in the Prepare
wizard.