Care
Essentials
CWS4000
Cocoon
Warming
System
Service
Manual
4.
PREVENTATIVE
MAINTENANCE
4.1.
Cleaning
e
Clean
the
CWS4000
control
panel,
enclosure
exterior,
and
hose
with
a
soft
cloth
lightly
dampened
with
a
non-staining
hospital
disinfectant.
e
Clean
accumulated
dirt
and
lint
from
the
air
inlet
slots
using
a
vacuum
cleaner.
4.2.
Electrical
Safety
Inspection
Care
Essentials
recommend
that
the
CWS4000
receive
regular
electrical
safety
inspections.
Information
on
the
type
and
frequency
of
inspections
may
be
obtained
from
locally
published
technical
standards.
In
Australia,
the
relevant
technical
standards
are:
e
AS3551
Technical
management
programs
for
medical
devices.
This
standard
specifies
procedures
required
to
develop
equipment
management
programs
for
medical
devices.
Some
of
these
include
procedures
for
acceptance,
fault
management
and
routine
testing
of
medical
devices.
This
standard
specifies
electrical
safety,
essential
safety
and
performance
testing.
e
AS2500
Guide
to
the
safe
use
of
electricity
in
patient
care.
This
standard
provides
a
comprehensive
guide
to
the
safe
use
of
electrically
operated
equipment
used
in
health
care
facilities.
Measures
are
detailed
to
provide
and
maintain
patient
and
operator
safety,
including
details
of
the
classes
of
equipment
and
electrical
installations
to
be
used
for
particular
medical
procedures.
Programmed
electrical
safety
inspections
are
essential
to
confirm
continued
operator
and
patient
safety.
Mandatory,
statutory
requirements
for
electrical
safety
inspections
may
also
apply.
Revision
|