6-2
Section 6 Troubleshooting
Cole Palmer
Low Temperature
adjustable high temperature protection limit
exceeded
check limit setting
check uid selection
Motor Fault
it can take over 10 minutes for the motor
temperature to get low enough before the unit
can be restarted
MOL
Motor Overload
allow unit to cool down
Open RTD1
Internal
open internal temperature sensor
Open RTD2
External
open external temperature sensor
Shorted RTD1
Internal
shorted internal temperature sensor
Shorted RTD2
External
shorted external temperature sensor
Bad Calibration
redo calibration
High Temperature
adjustable high temperature protection limit
exceeded
check limit setting
check uid selection
High Temperature
adjustable high temperature protection limit
exceeded
check limit setting
check uid selection
Low Level
reservoir uid level too low for safe operation
check uid level
Low Temperature
adjustable low temperature protection limit
exceeded
check limit setting
check uid selection
Warning Displays
The unit will continue to run with a warning. A warning also sounds the
alarm, if enabled.