Section 3 Installation
Ambient Temperature Range 10°C to 40°C (50°F to 104°F)
Maximum Relative Humidity 80% at 31°C (88°F)
Operating Altitude Sea Level to 2000 meters (6560 feet)
Overvoltage Category II
Pollution Degree 2
Degree of Protection IP 20
The unit is designed for continuous operation and for indoor use.
Never place the unit in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Refrigerated units should be left in an upright position for 24 hours at
room temperature before starting. This will ensure the lubrication oil
has drained back into the compressor.
Ambient
Conditions
Ventilation
The unit can operate with 0 clearance on two exhaust sides as long as the
third exhaust side has unrestricted air ow. Blocked ventilation will increase
the unit's temperature, reduce its cooling capacity and, on refrigerated units,
eventually lead to premature compressor failure.
Ventilation Options