Section 3 Installation
3-7
Cole Palmer
Filling Requirements
Draining
Ensure the reservoir drain port on the front of the unit is closed and that
all plumbing connections are secure. Also ensure any residue is thoroughly
removed before lling the unit.
Before using any uid refer to the manufacturer’s MSDS and EC
safety data sheets for handling precautions.
To avoid spilling, place your containers into the bath before lling.
With a low level WARNING the unit continues to run, with a FAULT the
unit will shut the refrigeration, pump and heater will shut down, see Section
7. The low level warning is at approximately 4 cm (1 5/8”) below the top, the
low level fault is at approximately 5.5 cm (2 1/8”).
Avoid overlling, oil-based uids expand when heated.
When pumping to an external system, keep extra uid on hand to maintain
the proper level in the circulating lines and the external system.
NOTE Monitor the uid level whenever heating the uid.
Before draining any uid refer to the manufacturer’s MSDS and EC
safety data sheets for handling precautions.
Ensure the uid is at a safe handling temperature, ~55°C. Wear
protective clothing and gloves.
place a suitable vessel underneath the drain. If desired, attach an •
8 mm id tube on the drain.
slowly
• turn the drain plug until ow is observed.
Additional Fluid
Precautions
When working with uids other than water:
Do not use any uid until you have read and understood the label and •
the Material Safety Data Sheet (MSDS).
Do not blend any uids. •
Ensure any uid residue or any other material is thoroughly removed •
before lling the unit with a different uid.
Always wear protective clothing, especially a face shield and gloves. •
Avoid spattering on any of the unit's components, always • slowly add uid.
When adding, point the opening of a container away from yourself.
Use fume hoods. •
Do not allow any ignition sources in the vicinity. •