COMMUNICATION MODE
The CryoWiz™ has the built in capability to communicate externally via the Ethernet (Web Server) or USB port.
The USB port is to be used only by Factory Authorized Technicians to customize the setup of the CryoWiz™.
The Ethernet port (Web Server) is used for all end user remote communication. This includes email notification,
system status viewing, and event logging.
CONNECTING THE WEB SERVER
The CryoWiz™ has an integral Web Server device. Connection to the Web Server is through an RJ45 Ethernet
connector located on the left side of the CryoWiz™. Unscrew the protective cap that covers the connector jack to
insert the Ethernet cable. Use any CAT 5 or CAT 6 cable to make this connection to the local area network. Refer
to the Web Server ADI included with the CryoWiz™ for instructions on interfacing it to the local area network
(CONCOA instruction manual 99065774 [ADI5774]).
POWER REQUIREMENTS
Input Voltage: 90 - 264 VAC, 50 / 60hz
Power Consumption: 20 watts
Date and Time Event Clock: 3 AA batteries.
SERVICE
For Service to the CryoWiz™, contact your local Distributor or CONCOA Customer Service at 1-800-225-0473.
FACTORY DEFAULT SETTINGS
Communication: USB, Ethernet (Web Server)
Units of Measure: °C, psig
Keypad Lockout Security: Disabled
Alarm Configuration: 5 Independent Alarm Relays
(Left Bank Depleted, Right Bank Depleted, Dry N2 Supply Pressure Low,
Oxygen Deficiency, Frozen Valve Alarm)
Audible Alarm: Enabled