17
3.3 Troubleshooting
Table 2: Troubleshooting Guide
Symptom Possible Cause Corrective Action
Handpiece does not
operate.
Handpiece cord not
connected securely.
Securely connect
handpiece cord to
controller.
Handpiece faulty. Return for service.
Error Messages
“MOTOR OVER
SPEED”
Handpiece and/or
System faulty.
Replace handpiece. If
symptom is corrected,
return handpiece for
service.
“HP SERVICE
REQUIRED”
Else, return system for
service.
“CANNOT ID
HANDPIECE”
messages display.
CONNECT
HANDPIECE
Handpiece not
connected securely.
Securely connect
handpiece to controller.
message displays when
the handpiece is already
plugged into the
console.
Handpiece and/or
System faulty.
Replace handpiece. If
symptom is corrected,
return handpiece for
service.
Else, return system for
service.