4.0 TECHNICAL SPECIFICATIONS
Medical electrical equipment complies with and was tested with respect to
electric shock, fire, electromagnetic compatibility, mechanical and other
specified hazards only, in accordance with UL60601-1, CAN/CSA C22.2 No.
601.1-M90, IEC60601-1:1988 +A1:1991 +A2:1995, ES60601-1:2005 +A1:
2009 +A2: 2010, CAN/CSA C22.2 No.60601-1-1-08 and IEC60601-1:2005
+C1:2006 +C2:2009.
Tested to IEC60601-1-2:2007 and Part 15 of the FCC Rules as follows: 1)
The system may not cause harmful interference: 2) The system will accept
interference, including interference that might cause undesired operation. If
interference occurs, separate the instruments. For more information, contact
customer service.
4.1 Product Technical Specifications
4.1.1 Handpieces
MC9840
Classification: Class I (intended for connection to Class I surgical
console).
Type rating of applied part is dependent on surgical
console.
Protection Ingress of
Fluids:
IPX0 (ordinary)
Mode of Operation: Intermittent Loading
Operating Speed
(Nominal):
300 - 5,000 rpm, Forward/Reverse
300 - 2,500 cpm, Single and Multi-turn Oscillate
Torque (Nominal): 19 in. oz.
Length: 7.15 in. (18.2 cm)
Diameter: Approx. 0.785 in. tear drop (2.0 cm)
Weight: 17.3 oz. (490g)
Cord length: 10 ft. (3m) Non-detachable
Material: Anodized Aluminum and Stainless Steel
Duty Cycle (once daily)
with fluid flow:
30 seconds ON, 15 seconds OFF (3x)
3 minutes ON, 30 seconds OFF (4x)