3
6. Do not excessively bend or kink the handpiece cord. Always inspect
cords for signs of excessive wear or damage, or bent, broken or missing
pins within the connector(s). If any wear or damage is found, discontinue
use and replace handpiece immediately. Using a damaged power cord
could possibly cause injury.
7. Handpieces are supplied non-sterile. Clean and sterilize prior to each
use.
8. Do not contact the moving parts on the handpieces. Injury to the
operator may occur.
9. Continually check handpiece for overheating. If overheating is sensed,
immediately discontinue use and return equipment for service. Overheating
of the blade or bur may cause damage to the blade or bur and may cause
burns or thermal necrosis.
10. While handpiece is not in use do not place on patient/surgical drapes. Place
handpiece on mayo stand.
11. Do not immerse the equipment in fluids. Immersion may render the
device inoperable.
12. Failure to follow the specified service interval could result in reduced
instrument performance or overheating of the handpiece.
Overheating can lead to possible burn injury to the patient or medical
personnel. Rotation of handpiece usage per day will assist with proper
performance. (Refer to section “3.4 Maintenance Schedule”).
13. Do not attach, insert or remove accessories or attachments while the
handpiece is operating. Injury to the operator and/or damage to the
equipment may occur. Place the handpiece safety mechanism in the “safe”
position prior to installation or removal of items.
14. Avoid contact with cutting tip of blade or bur when locking into handpiece.
Tips are sharp and may cause injury.
15. After use blades, burs and tubing sets may be a potential biohazard
and should be handled and disposed in accordance with acceptable
medical practice and applicable local and national requirements.
16. Do not use burs for plunge cutting. Injury or damage may occur.
!