Telehealth medical diagnostic system IDIS2GO
Picture 21: Time settings interface Picture 22: Minute setting interface
d) Calibration
In theSettingsinterface select “Time” to enter theCalibrate parameterinterface, as shown in the
Picture 23. Select 2L or 3L according to the volume of the syringe, then go to the “Calibration” interface, as shown
in the Picture 24.
Picture 23: Calibration interface Picture 24: Calibration interface
Under theCalibration, interface, press the syringe once, the device will show "REPEAT", then press the
syringe again. After two correct continuous operations, the calibration will be successful and the device will show
“OK”. Finally, the interface will switch to the previous interface before calibration (previous interface: if the device
is calibrated after the measurement is completed, it will return to the Settings interface; if it is calibrated
before the measurement is completed, it will return to the Examination interface).
If “Error! Please repeat” is displayed on the device, it means something is wrong with the operation, please
repeat the calibration until success. If “Select the correct volume” appears on the device, please confirm the syringe
volume and calibration selection, and then repeat the calibration until successful. If you need to stop the calibration,
just press the “Confirm” button to go to the previous interface before calibration.
In theCalibrate parameterinterface select “Adjust” to enter theAdjustinterface, as shown in the
Picture 25. Press the Up or Down buttons to change the value (long press available), then press the Confirm button
to return to theConfirm adjustmentinterface, as shown in the Picture 26. Selecting “Yes” will save the
configured value, selecting “No” will cancel the setting, and the device will return to theCalibrationinterface.
Notation
The set values determine the accuracy of the measurement, please DO NOT change them randomly. After
replacing the turbine, calibration is used to enter the parameters of the new turbine, which guarantees the
accuracy of measurements after replacing the turbine.
Picture 25: Calibration interface Picture 26: Adjustment confirmation interface
In the Calibrationinterface select “Exit” to return to theSettingsinterface.
e) About
In theSettingsinterface select “About” to enter theAboutinterface. The user can view the device
name and software version. Click the "Confirm" button to return to theSettingsinterface.
f) Exit
In theSettingsinterface select “Exit” to return to theMenuinterface.
Power off
In the Menuinterface select “Power off”, the device will turn off.
Notation. If the operation is not taken within 1 minute, the device will turn off automatically.
Exit