EasyManua.ls Logo

Controlotron 1010PVDN - 2.2 Flowmeter Setup; Flowmeter Configuration Steps

Default Icon
203 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Addendum 1010FMA-4
2.2 Flowmeter Setup
NOTE: The following instructions require the use of the keypad and the menu.
The installer should become familiar with their use before proceeding
further.
2.2 . 1 Switch the flowmeter on. Press the <MENU> key.
2.2 . 2 On multi-channel flowmeters, use the arrow keys to select [Dual Channel Flow]
or [Dual Beam Flow] depending on the mode utilized for the wet-flow calibra-
tion.
2.2 . 3 Use the arrow keys to select either [Clamp-on], [Flow Tube] or [Clamp-on Spool].
2.2 . 4 Select [Full Site Setup] and use the <Right Arrow> to select [Channel Setup];
then select [Recall Site Setup].
2.2 . 5 Use the <Down Arrow> to scroll to the site name indicated on the Calibration
Diagnostic Data Sheet. Then press <ENT>.
2.2 . 6 The meter will perform a momentary Makeup routine that will take a few
seconds and then begin operation.
2.2 . 7 Refer to your 1010 field manual for instructions on setting zero flow.
NOTE: Setting zero flow must be performed each time the transducers are in-
stalled. The zero adjustment has no effect on the wet-flow calibration
data or the calibration (Kc) factor.
2.2 . 8 Using the arrow keys, scroll to the Data Span/Set/Cal menu location. Verify that
the [Kc] (calibration) factor matches the value indicated on the diagnostic data
sheet.
2.2 . 9 If you are measuring a liquid other than ambient water, select the [Liquid Class]
menu cell and <Down Arrow> from there to [Viscosity]. Enter the correct vis-
cosity for the liquid you will be monitoring.
2.2 . 10 Setup is now complete. Press the <MENU> key twice to view the flow rate dis-
play. DO NOT utilize the [Save Site] command when it appears.
2.2 . 11 When measurements are completed, simply turn off the meter. DO NOT save
the site. This might contaminate the wet-flow calibration data already stored.
NOTE: Contact Controlotron’s Technical Services Department if any flow cali-
bration data is accidentally removed or overwritten.
2

Table of Contents

Related product manuals