Prole Programming
Page 29 CXX-UM-03060-EN-01April 2019
Pressure Calibration
Figure 21: Pressure tab
The Pressure table is used for calibrating pressure transducers. Pressure calibration
does not apply to the EC80 flow processor.
Temperature Compensation
Figure 22: Temperature tab
The Temperature table provides the necessary calibration of the flow
processor’s temperature input device. Obtain a temperature profile by testing
the temperature sensor being used. The temperature profile is unique to the
individual sensor. To calibrate the temperature sensor:
1. Place the temperature sensor in a temperature bath.
2. Enter the temperature vs. counts data into the Temperature
Compensation table.
Typical applications use either a 10k Thermistor or 100 Ohm Platinum
RTD. This provides the flow processor the look-up table necessary for
temperature acquisition.