Conguring Liquid Properties
Page 35 CXX-UM-03060-EN-01April 2019
The software calculates the viscosity across the entire specified temperature
range. The most commonly used and most accurate method for calculating
viscosity is the ASTM method. However, the Andrade method may be used.
OTE:N All calculations are done per ASTM D341 when ASTM is selected.
Temperature vs Density Table
Figure 29: Temperature vs Density Table
The density table is only applicable when a mass flow rate is needed. Using the
flow meters temperature sensor, density is calculated from the Temperature vs
Density table and then multiplied by the volumetric flow rate to obtain the mass
flow rate.
MassFlowRate = VolumetricFlowRate × Density
Leave these fields blank or clear the table using the IDX button (right-click the IDX
button and select Clear Grid), if a mass flow rate is not needed.
You will need to get the density properties from our factory or independently
in order to calculate mass flow rate. If difficulty arises in obtaining density
information regarding your liquid, have Badger Meter test the liquid properties in
our laboratory, and supply you with the temperature vs density information.
Liquid Pressure Compensation
Figure 30: Liquid Pressure Correction Coefficient table
Liquid pressure compensation is not available for the EC80 Flow Processor.