Flow processor Programming
April 2019CXX-UM-03060-EN-01Page 40
Option Function
Manual Stop
Manual stop begins data logging when you click Start and ends data logging when
you click Stop.
Timed Stop
(Seconds)
Timed stop (Seconds) begins data logging when you click Start and ends data logging
when it reaches the numerical value adjacent to the drop-down menu. For Example,
if 100 were placed in the time field adjacent to the drop-down menu, 100 seconds
would elapse before the data would cease to log.
Timed Stop
(Minutes)
Timed stop (Minutes) begins data logging when you click Start and ends data logging
when upon reaching the numerical value adjacent to the drop-down menu. For
Example, if 100 were placed in the time field adjacent to the drop-down menu, 100
minutes would elapse before the data would cease to log.
Timed Stop
(Date)
Timed stop (Date) begins data logging when you click Start and ends data logging
when upon reaching the numerical value adjacent to the drop-down menu. For
Example, if 1/1/20 were placed in the time field adjacent to the drop-down menu, the
GUI would continue to log data until January 1st 2020.
Sample Stop
Sample stop begins data logging when you click Start. The software logs one sample's
worth of data and stops at the set interval in the sample rate field.
Viewing Log
To view the log being populated, click View. You will be transferred to the logging
file you have designated. Values update at the intervals set.
Watching Log
To watch the log being populated in real-time, click Watch. You will be transferred
to the logging file you have designated.
FLOW PROCESSOR PROGRAMMING
Connect all sensors, transducers, pickups and com-link cables to the EC80 Flow
Processor before powering up the computer. Communication needs to be
established with the EC80 Flow Processor prior to uploading or downloading
a profile. All flow processors, unless specially configured, communicate via an
RS485 connection to the PC. The PC, in conjunction with an RS485 converter is
required for communicating with the flow processor. Ask your sales representative
regarding purchase of USB-to-RS485 converters (PN EF I-7561).
Downloading Proles
Downloading a profile is how you use all the information you programmed into
the IFC GUI to program the flow processor. Follow this procedure to download the
current profile to the flow processor:
1. Click the white box next to Enable on the tool bar.
2. Click the Download icon . All information in the data tables, either blank or
populated, is written to the ow processor.