3C Interfacing and Programming CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-16 CMP 200
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.5.2 Error Log (Cont)
Step Action (membrane console)
Action (PC GenWare) Action (Touchscreen GenWare)
2.
Select ERROR # and use the + or -
buttons to scroll through the error log.
The error code, error message, date and
time of the error will be displayed in the
LCD window, and the associated
parameters will be displayed on the
console displays
Click on the
< or > buttons on the Error
Log window to scroll through the error
log.
The error code, error message, date and
time of the error will be displayed on the
left side of the Error Log window, and the
associated parameters will be displayed
under Rad Parameters.
Press the
< or > buttons on the Error Log
window to scroll through the error log.
The error message will be displayed to
the right of the
< > buttons on the Error
Log window. The error code, date and
time of the error, etc. will be displayed
under Error Details on the left side of the
Error Log window, and the associated
parameters will be displayed under
Radiographic Parameters.
3.
Press EXIT to return to the UTILITY
menu.
3C.5.3 Statistics
This utility shows the tube exposure count and the accumulated generator exposure count. This also allows resetting of the tube 1 exposure
counter.
The STATISTICS menu for the membrane console is shown below.
* STATISTICS *
TUBE 1 EXP: 500 RESET TUBE 1 EXP
TOTAL EXP: 1100
EXIT