Cygnus 2+ Operating Manual
Cygnus M5-C2P
Technical Specifications
(in steel)
Single element 0° probes in Multiple Echo mode;
Twin element probes in Single Echo mode;
Twin element probes in Echo-Echo mode;
Twin element probes are zeroed in the air without the need of
a zero block. The user is prompted to zero the probe at power
up and after a probe change.
Single element probes in multiple echo mode do not need to
Single Echo mode with Twin element probe. Measures
thickness using the time from the delay line (zero point) to the
first back-wall echo. (UT Mode 2)
Echo-Echo mode with Twin element probe. Measures thickness
using the time between the first and second back-wall echoes.
(UT Mode 3)
Multiple Echo using Single element probe. Uses three matched
and verified back-wall echoes to determine the material
thickness and ignore surface coatings.
High Temperature
Measurement
All standard single element and twin element probes:
Allows continuous measurement up to 75°C (160°F).
Single Echo and Echo-Echo measurement modes;
Multiple Echo measurement modes;
Low
(measurement > 120 mm)
Medium
(measurement < 120 mm)
±0.1 mm (±0.004”) or 0.1% of thickness measurement whichever is