Cygnus 2+ Operating Manual
Cygnus M5-C2P
Technical Specifications
128 x 32 Pixels. 25.58 mm (W) x 6.38 mm (H)
Digital Thickness Value. Settings. Battery Level.
Pulse Energy : Voltage (peak-
to-peak)
Pulse Energy : Pulse Duration
S2D : 220 ns
S3C : 100 ns
S5A / S5C : 67ns
T2C : 220 ns
T5B : 100 ns
T7A : 67 ns
Automatic Gain Control depending on probe and measurement
mode.
1.0 MHz to 10.0 MHz (-6dB)
USB Mini B Connector under battery cover.
Can be used to update the gauge firmware and load gauge
Calibration setting storage
Calibration Data stored to Internal Flash Memory
Not required for Multiple Echo mode.
Automatic V-path correction for twin element probes.
Option of two point calibration for twin element probes.
Display & Recall Facilities
IPX67 (Water immersion 1 metre depth for 30 minutes)
MIL STD 810G Method 501.6 (High Temp +55°C)
MIL STD 810G Method 502.6 (Low temp -20°C)
MIL STD 810G Method 507.6 (Humidity 95%)
MIL STD 810G Method 512.6 (Immersion 1m, 30min)
MIL STD 810G Method 514.7 (Vibration)
MIL STD 810G Method 516.7 (Shock 20g)
MIL STD 810G Method 516.7 (Transit Drop 1.22m)
Designed for BS EN 15317:2000.