23
8-5. Calibratuib by Shunt CAL Value
It is to calibrate with the S.CAL value written down for load calibration. You can
calibrate without any standard weight.
Measuring mode
While key is being Press, press key
Display Function mode
Display Sensor output value calibration
Display Actual load calibration indicate
Input the rated output value by using and key. And
then. Press key
Calibration end
Display Shunt cal value calibration mode
Display R.O. value input mode
Display Shunt cal value input mode
Input Shunt cal value by using and key. And then.
Press key