EasyManua.ls Logo

dacell DN-130L - Actual load calibration

Default Icon
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
8-4. Actual load calibration
This is a calibration method by adding actual load on the sensor. Standard weight is needed
.
Measuring
mode
Display Function mode
Display Sensor output value calibration mode
Display Actual load calibration mode
Display actual load mode
Input the rated output value by using and key.
And then. Press key
Calibration end
Display zero mode. press key without any
load on the sensor
About 1 sec later
After adding the actual load (standard weight) on the sensor,
press key.
About 1 sec later
Display R.O. value input mode
Input the rated output value by using and key. And then.
Press key (Input 3.0000 if do not know the R.O. value)
Display Shunt CAL value (please write to S.CAL)
Channel R.O. value, calibration constant
CH1
CH2
While key is being Press, press key
Select channel 1 or 2