8-4. Actual load calibration
This is a calibration method by adding actual load on the sensor. Standard weight is needed
.
Display Sensor output value calibration mode
Display Actual load calibration mode
Input the rated output value by using and key.
And then. Press key
Display zero mode. press key without any
load on the sensor
After adding the actual load (standard weight) on the sensor,
press key.
Display R.O. value input mode
Input the rated output value by using and key. And then.
Press key (Input 3.0000 if do not know the R.O. value)
Display Shunt CAL value (please write to S.CAL)
Channel R.O. value, calibration constant
CH1
CH2
While key is being Press, press key