ENGLISH | 13
Readings and normalization
of the volumes
e totalizer of the installed dry gas meter has a
resolution of 0.1 liters.
Before starting the sampling, take note of the vol-
ume and temperature values.
At the end of the operation, read again the volume
indicated and note it, with temperature value as
well.
Normalization of the volume can be done using the
following formula:
Vs = Vm .
Pa
.
Tn
101.3 273 + Tgm
Where:
Vs [lt] = Normalized volume
Vm [lt] = Dry gas meter total volume
Pa [kPa] = Ambient pressure
Tn [K] = Normalization temperature
Tgm [°C]= Dry gas meter temperature