20 | QB1 User Manual
Sampled gs conditions dry with max temperature 45°C
Gas inlet With protection lter and quick connector
Working temperature -10 ÷ 40°C 95% UR
Stocking temperature -10 ÷ 50°C 95% UR
Construction Inox steel and aluminium
Size 330 x 310 x 360 mm (L x P x H)
Characteristics and accuracy of the measures
Volume
Dry gas meter class G1.6
Flowrate range 0.016 m3/h ÷ 2.5 m3/h
Accuracy 2% of the measure
Resolution 0.1 liters
Ball ow meter
Range see the available models table
Accuracy 5% f.s.
Dry gas meter temperature
Range -50 ÷ 70°C
Accuracy ± 1 °C
Resolution 0.1 °C
Suction pressure drop (excluding V2x5DC model)
Vacuum meter Range 0 ÷ 1 Bar
Accuracy ± 5%