6 Datacolor 200 User Guide
Controls and Indicators
Status Panel
The status panel indicates the current instrument settings being used. Below is the status panel for the
Datacolor® 200:
OPTION
Calibration. Indicates if the calibration interval has expired.
Red. Calibration has elapsed. Recalibrate the
instrument before using to ensure an accurate
measurement.
Green. Instrument is calibrated.
Service. Indicates when the instrument should be serviced.
Amber. Annual service check is recommended.
Green. Instrument is performing properly.
Powering Up
WARNING
The power cord supplied with the unit must be used.
1. Verify all cable connections.
2. Plug the AC power cord into an outlet.
3. Attach the instrument cable to the instrument, and
to a port on the computer
4. Turn on the computer.
5. Turn on the power switch. It is the red switch on
the back of the instrument.
6. When power is applied, all mechanisms are
automatically reset.
NOTE
If the instrument is not turned on before you launch a Datacolor program, you may receive an error
message.
1.
When the front panel Calibration LED remains lighted, your instrument is ready to be calibrated.
2.
Launch a Datacolor program.
Power switch