Datacolor 200 User's Guide 13
Sample Presentation and Measurement
When measuring samples using Datacolor® 200, you must pay close attention to the positioning of the
sample to insure an accurate measurement.
When positioned correctly, the sample rests between the
sample holder and the front panel door. The sample must completely cover the aperture opening.
To position the sample:
1. Grasp the sample holder and pull forward.
2. Position the sample at the port.
3. Return the sample holder to the upright position.
WARNING
Do not allow the sample arm to spring back.