WP
4
PotentiaMeter
Calibration and Verification Standards
30
6. At this screen, adjust the water potential value to
its proper value for the KCl solution at that tem-
perature (-2.19 MPa at 20 °C; -2.22 MPa at 25
°C). Press the upper right button to move the
value up, the lower right button to move it
down. When it is at the value you want, press
the Exit button. The value will be stored.
Note: This is the only menu where these buttons can
change the calibration, so you won’t hurt anything
by pressing these buttons in other menus.
7. Read the KCl standard again in the normal sam-
pling mode. It should read the proper value.
8. If after adjusting the calibration and cleaning the
chamber you still are getting incorrect readings
when reading the KCl, contact Decagon at 509
332-2756 (1-800-755-2751 in US and Canada) for
further instructions.