WP
4
PotentiaMeter
Cleaning and Maintenance
48
10. Cleaning and Maintenance
The accuracy of your WP4 is vitally dependent on
keeping your instrument clean. Dust and sampling
debris can contaminate the sampling chamber and
must therefore be regularly cleaned out. To clean
your instrument, carefully follow these instructions.
Tools Needed
thin plastic rod or other non-metal implement
Distilled Water
Isopropyl Alcohol
Lint-free or sizing-free tissues (Kimwipe
®
)
or
WP4 Cleaning Kit
Note: Kimwipes
®
are ideal because they don’t leave a
lint residue like most tissues. They also don’t have
any other compounds in the tissue that may contam-
inate the sensors in the WP4s block. Also, never use