PotentiaMeter
Cleaning and Maintenance
51
4. Wrap a new dry Kimwipe around the spatula,
and use it to thoroughly wipe the cleaning solu-
tion from the mirror surface and chamber.
Optical Sensor:
You will probably clean the optical sensor while
you are cleaning the mirror, since they face each
other in a very small gap. Clean it in the same man-
ner as described above for the mirror.
Thermopile Sensor:
Wrap a strip of Kimwipe tissue around a plastic
cleaning kit spatula and moisten it with water or
alcohol to clean the thermopile sensor. This sensor
must be free of all dirt and lint.
Inside Chamber:
Clean all other portions of the chamber block with a
moist Kimwipe.
Note: Be extremely careful not to damage the fan
blades in the chamber. The fan blades are very frag-
ile; if one of them breaks, your instrument won’t
work properly. Therefore, take extra care when
cleaning this portion.
Inside Case:
1. Clean the sample drawer and drawer base.
Remove any debris that may have collected