WP
4
PotentiaMeter
Troubleshooting
62
machine:
SOLUTIONS:
• The block is not plugged in to the motherboard.
Open the case and check to make sure that the
small ribbon cable that connects the block to the
motherboard is snapped and locked in place.
• One or more components has failed on the
block’s circuit board. If the block is properly
plugged in to the motherboard and this message
appears, it is likely that one or more of the com-
ponents have failed on the block’s circuit board.
If you press
-exit-
at this prompt, the instru-
ment will not have values for the component
that has failed, which will lead to incorrect read-
ings. If this message appears and you continue
to sample, Decagon cannot be liable for errors
in reading that may occur. Contact Decagon for
a solution to this problem.
7. PROBLEM:
“TE Set” option does not appear anymore in the
WP4-T’s system configuration menu.
Block ROM failed
-exit- see Manual