For more information about the Flow Control and Cleaning Frequency, see 21.2.2
Functional Button “ON”.
10.4 System Settings
At commissioning, alarms set points and alarm delays can be set. The signal output can
be modified from 0 20 mA or 4 20 mA. According to external requirements also the
offset can be set from -5 ppm to 5 ppm. Also, the settings can be reset to the factory
default values, if needed.
For more information about the Settings, see 21.2.3 Functional Buttons AL1/ AL2/
SET.
10.5 Auto Clean Control Unit
Figure 23. Auto Clean Control Unit
A.
Air Regulator
D.
Air Flow Adjuster Tube
B.
Air Flow Adjuster
E.
Wiper Piston
C.
Air Regulator Tube