Page: 57
Status
Com?
Sampe?
Data log?
Reading
EE
0..49/EE
System-Alarm-LED
Red/Steady
Alarm Circuit 1,2
Alarm
Reason
No communication
between Computer
Unit and Measuring
Cell.
Meter is not able to
measure the sample:
no water in oil
content much too
high no light
transmission
possible.
Data logging is not
possible:
No DECKMA
memory card
inserted.
A read only card has
been inserted.
-A new DECKMA
memory card has
been inserted, but
has not been
activated.
Recommendations
Check connection
between Computer
Unit and Measuring
Cell.
-Check sample,
check sample flow
rate, clean Sample
Glass Tube
according to
section11. OMD-
24 Series Operating
instructions
Insert the active
memory card.
New memory card.
Activate the
memory card.
Status
Int.Err
Desicc.
Humid
Reading
0..49/EE
System-Alarm-LED
Red/Steady
Green/ Blinking
Alarm Circuit 1,2
Alarm
Normal Operation
Reason
Internal error.
Measuring Cell
Humidity critically
high (>40%rH)
Sample temperature
below dew point.
Instantaneous
condensation
possible.
Recommendations
Restart the system.
Check/ Replace
desiccator.
Check/ Replace
desiccator.