26
DAC-510E-EN-BF
“Stop now”, “Do not touch the patient”–Thisindicatesthatthetwo-minuteCPRperiodhas
endedandtheusershouldstopCPR.TheunitwillenterAnalyzingmodeandthe“analyzing”LEDwill
flash.
4.8 Post-Shock CPR
IftheDDU-100AEDhasdeliveredashock,theunitwillrequireamandatorytwo-minuteCPRperiod.
No patient ECG rhythm monitoring will be done during this period. Once the two-minute period is
complete,theAEDwillcontinueinAnalyzingmode.
Post-Shock CPR Voice Prompts:
“It is safe to touch the patient”This indicates that it is safe for the user to touch the patient. The unit
willnotbemonitoringthepatient’sECGrhythmduringthisrequiredtwominuteCPRperiod.The “do not
touch patient” LED will be off to indicate that it is safe to touch the patient.
“Begin CPR now”–ThisindicatesthattheusershouldperformCPRfortwominutes.Theunitwillnotbe
monitoringthepatient’sECGrhythmduringthisrequiredtwo-minuteCPRperiod.The“analyzing”LEDwill
remain off to indicate that background rhythm monitoring has been suspended.
“Continue for ‘x’ seconds” or “Continue for 1 minute ‘x’ seconds” This indicates that the user
shouldcontinueperformingCPRfor‘x’moresecondsor,for1minuteand‘x’moreseconds,respectively.
Theunitwillnotbemonitoringthepatient’sECGrhythmduringthisrequiredtwo-minuteCPRperiod.The
“analyzing”LEDwillremainofftoindicatethatbackgroundrhythmmonitoringhasbeensuspended.
“Continue”–ThisindicatesthattheusershouldcontinueperformingCPR.Thisphraseisspokenbetween
the“continuefor‘x’seconds”or“continuefor1minute‘x’seconds”promptstolettheuserknowthatthe
unitisstilloperatingnormally.Theunitwillnotbemonitoringthepatient’sECGrhythmduringthisrequired
two-minuteCPRperiod.The“analyzing”LEDwillremainofftoindicatethatbackgroundrhythmmonitoring
has been suspended.
“Continue for 5, 4, 3, 2, 1”, “Stop CPR” –ThisindicatesthattheusershouldnishperformingCPR.
Thisphraseisspokenduringthelastseveralsecondsoftherequiredtwo-minuteCPRperiodtolettheuser
know that the unit is still operating normally and the two-minute period is ending.
“Stop now”, “Do not touch the patient” –Thisindicatesthatthemandatorytwo-minuteCPRperiod
hasendedandtheusershouldstopCPR.TheunitwillenterAnalyzingmodeandthe“analyzing”LEDwill
flash.