53
DAC-510E-EN-BF
8.1.7 Guidance and Manufacturer’s Declaration – Electromagnetic Emissions and Immunity
Electromagnetic conformity
Guidanceandmanufacturer’sdeclaration-TheDDU-100isintendedforuseinthe
electromagnetic environment specified below. The customer or the user of the DDU-100 should
assure that it is used in such an environment.
Electromagnetic emissions
Emissions test Compliance Electromagnetic environment – guidance
RFemissions
CISPR11
Group 1
Class B
TheDDU-100usesRFenergyonlyforitsinternal
function.Therefore,itsRFemissionsareverylow
and are not likely to cause any interference in nearby
electronic equipment.
The DDU-100 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable
Electromagnetic immunity
Immunity test IEC 60601 test
level
Compliance level Electromagnetic environment -
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
The DDU-100 uses interference
detected and motion detected
indicators to notify the user if
conditions are not ideal. No other
ESD requirements are necessary.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power line
supply lines
±1 kV for input/
output lines
Not applicable
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
Not applicable