47
DAC-510E-EN-BF
8 Technical Specifications
8.1 Defibtech DDU-100 AED
8.1.1 Physical
Category Specification
Size
8.5 x 11.8 x 2.7 inches (22 x 30 x 7 cm)
Weight
Approximately4.2lbs(1.9kg)withDBP-1400Batterypack
Approximately4.4lbs(2kg)withDBP-2800Batterypack
8.1.2 Environmental
Category Specification
Operating / Maintenance
Temperature 0 – 50°C (32 – 122°F)
Humidity 5% – 95% (non-condensing)
Standby / Storage
Temperature 0 – 50°C (32 – 122°F)
Humidity 5% – 95% (non-condensing)
Altitude
-150 to 4500 meters (-500 to 15,000 feet) per
MIL-STD-810F500.4ProcedureII
Shock / Drop Abuse Tolerance
MIL-STD-810F516.5ProcedureIV(1meter,anyedge,
corner, or surface, in standby mode)
Vibration
MIL-STD-810F 514.5 Category 20
RTCA/DO-160D,Section8.8.2,CatR,Zone2,CurveG
(Helicopter)
RTCA/DO-160D,Section8,CatH,Zone2,CurvesB&R
(Jet Aircraft)
Sealing
IEC60529classIP54;DustProtected,SplashProof,
(Battery pack installed)
ESD
EN 61000-4-2:1998 Severity Level 4
(Open air discharges up to 8 kV or direct contact
discharges up to 6 kV)
EMC (Emission)
EN 60601-1-2:2001+A1:2006, method EN 55011:1998
Group 1 Level B
(Nottoexceed30dBμVfrom30Hzto230MHzand
nottoexceed37dBμVfrom230to1000MHz
EMC (Immunity)
EN 60601-1-2:2001+A1:2006, method EN 61000-4-
3:1998 Level 3
(Field strength: 10V/m; carrier frequency range: 26
MHzto1GHz;AMmodulation,80percentindex,at3
frequencies:1,5,and20Hz)