46
DAC-510E-EN-BF
7.2 Downloading the Internal Data Log
RegardlessofwhetheraDDCisinstalledintheunit,selectinformationisrecordedinternallywithin
the DDU-100 AED. The information recorded is limited to:
• ThetimetheAEDwasturnedon.
Other data such as event milestones (motion detection, shock advice, shock delivery
information, etc).
Eight seconds of ECG data immediately before a shock/no-shock decision, eight seconds
immediately after each shock, and all ECG data during the charging and waiting-to-shock
periods.
Note: Audio data is not logged internally.
To download the internally logged information, perform the following procedure:
Insert a blank DDC into the unit.
Turn the unit on.
Once the unit is on, turn it off in data download mode by pushing and holding the ON/OFF
button for at least five seconds.
Allow the unit to write the contents of the internal log to the DDC by waiting for the unit to
turn off automatically.
The DDU-100 will write the contents of the internal log onto the DDC. This information can then be
reviewed using the DefibView software.