Patient Impedance
(Ohms)
Phase A, Duration
(msec)
Phase B, Duration
(msec)
Energy Delivered
(Joules)
25 2.8 2.8 153
50 4.1 4.1 151
75 7. 2 4.8 152
100 9.0 6.0 151
125 12.0 8.0 153
150 12.0 8.0 146
175 12.0 8.0 142
Pediatric
Patient Impedance
(Ohms)
Phase A, Duration
(msec)
Phase B, Duration
(msec)
Energy Delivered
(Joules)
25 4.1 4.1 35
50 5.8 3.8 47
75 5.8 3.8 51
100 7. 2 4.8 53
125 7. 2 4.8 52
150 9.0 6.0 53
175 9.0 6.0 51
8.1.5 Patient Analysis System
ThePatientAnalysisSystemensuresthatthepad/patientimpedanceiswithintheproperrange
andanalyzesthepatient’sECGrhythmtodeterminewhetherashockisrequired.Ondetectionof
anon-shockablerhythm,theuserispromptedtoperformCPR.Forshockablerhythms,theAED
automatically charges in preparation for shock delivery.
The patient analysis system will detect electrical “noise” or artifact in the ECG signal that may
interfere with accurate rhythm analysis. This artifact may be caused by excessive motion to the
patient or by external electrical noise. When this artifact is present, the AED will prompt the user to
“Stop motion” or “Stop Interference” until the ECG signal is free of noise and then proceed to analysis.