Immunity test IEC 60601 test
level
Compliance level Electromagnetic environment -
guidance
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Not applicable Not applicable
Powerfrequency(50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Powerfrequencymagneticelds
should not be greater than levels
characteristic of a typical location
in a commercial or hospital
environment.
RadiatedRF
IEC 61000-4-3
10 V/m
80MHzto2.5GHz
10 V/m PortableandmobileRF
communications equipment should
be used no closer to any part of
the DDU-100, including cables,
than necessary. The recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter is
shown in the following table.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Note1:At80MHzand800MHz,thehigherfrequencyrangeapplies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
TheISM(industrial,scienticandmedical)bandsbetween150kHzand80MHzare6,765MHzto6,795MHz;
13,553MHzto13,567;26,957MHzto27,283MHz;and40,66MHzto40,70MHz.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy.ToassesstheelectromagneticenvironmentduetoxedRFtransmitters,anelectromagneticsitesurvey
should be considered. If the measured field strength in the location in which the DDU-100 is used exceeds the
applicableRFcompliancelevelabove,theDDU-100shouldbeobservedtoverifynormaloperation.Ifabnormal
performance is observed additional measures may be necessary, such as reorienting or relocating the DDU-100.