37
Trigno Recharging Base Station
(1) Range is characterized in normal office environments. Interfering RF sources in the 2.4GHz
spectrum, as well as absorptive objects occluding the RF communication path may degrade
transmission distance. Stated range can be exceeded under favorable RF conditions.
(2) All data packets are guaranteed to by synchronized within the established sensor network.
(1) Delay from sensor event to analog output is fixed for all channels, and referenced as a
"group delay". In the case of EMG channels, this delays consists of the following elements: a)
Sensor Filter (1.5 ms), b) Digitization Delay (41 ms), c) SinX/X Correction (2 ms), and d)
Analog Low Pass Filtering (3.5 ms).
(2) This parameter relates the detected voltage input at the sensor to the reproduced voltage at
the EMG channel output.
(3) Sin(X)/X correction is applied to each channel for cancelling the distorting effects of the DAC
sampling process.
GENERAL SPECIFICATIONS
Typical Operating Range
(1)
40 m
RF Frequency Band 2400-2480 MHz (ISM)
Power Consumption
<6 W (operation)
<14W (during recharge)
Sensor Recharge Time < 2.5 hours
USB type USB 2.0 compliant, high speed
Temperature Range 5 - 50 degrees Celsius
Maximum number of Sensors 16
Inter-sensor delay
(2)
0 seconds
Dimensions 276 x 241 x 127 mm
EMG ANALOG OUTPUT SPECIFICATIONS
Number of outputs 16
Signal group delay
(1)
48 ms
EMG Signal Range ±5 V
Effective EMG Signal Gain
(2)
909 V/V ±5%
Channel Offset (r.t.o.) ±100 mV (max)
Baseline Noise <0.5mV RMS
DAC Filter Bandwidth DC-500 Hz, 160 dB/Dec.
Passband Ripple
(3)
<2% with Sin(X)/X correction
Connector Type SCSI-68, Type II