EasyManua.ls Logo

DeVilbiss 7305 SERIES - Approvals and Classifications; Product Approvals; Equipment Classifications; ISO Classifications

DeVilbiss 7305 SERIES
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7305D Series Two-years limited, excluding container
Internal Battery 90-day
Approvals
7305D & 7305P Series IEC 601-1; CAN/CSA-C22.2 No. 601.1-M90; UL 60601-1, EN 60601-1-2
7305P meets RTCA/D-160D DO-160D – section 21 Category M For battery operation only Airline use
Equipment Classifications
With respect to protection from electric shock Class I and internally powered
Degree of protection against electric shock Type BF Applied Parts
Degree of protection against ingress of liquids IPX2 and ordinary power supply
Mode of Operation Intermittent Operation: 30 minutes on, 30 minutes off
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
ISO Classification
7305P Series only – Electrically powered medical suction equipment for field and transport use according to ISO 10079-1 : 1999
High Flow/High Vacuum
7305D Series – Electrically powered medical suction equipment for non-transport use according to ISO 10079-1 : 1999
PROVIDER’S NOTES
No routine calibration or service is required provided the device is used in accordance with the manufacturer’s directions. In case of a change of patient, the device must be reconditioned to
protect the user. Reconditioning must only be carried out by the manufacturer or service provider. Between patients:
1. Visually inspect unit for any damage, missing parts
2. Ensure that unit and accessories are
3. Use an independent vacuum gauge to verify the unit provides the proper vacuum level as stated in
4. Discard and replace collection container, filter, and tubing between
5. Wipe the surface using a clean cloth dampened with
TWO-YEAR LIMITED WARRANTY
The compressor portion of the DeVilbiss Suction Unit 7305P & 7305D Series (excluding internal rechargeable batteries) is warranted to be free from defective workmanship and materials
for a period of two years from date of purchase. Internal rechargeable batteries are warranted for 90 days. Any defective part(s) will be repaired or replaced at DeVilbiss Healthcare’s option
if the unit has not been tampered with or used improperly during that period. Make certain that any malfunction is not due to inadequate cleaning or failure to follow the instructions. If repair
is necessary, contact your DeVilbiss Healthcare Provider or DeVilbiss Service Department for instructions: U.S.A. 800-338-1988 or 814-443-4881, Europe +49-(0) 621-178-98-0.
NOTE–This warranty does not cover providing a loaner unit, compensating for costs incurred in rental while said unit is under repair, or costs for labor incurred in repairing or replacing
defective part(s).
THERE IS NO OTHER EXPRESS WARRANTY. IMPLIED WARRANTIES, INCLUDING THOSE OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED
TO THE DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED. THIS IS THE
EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION
IS PERMITTED BY LAW. SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, OR THE LIMITATION OR EXCLUSION OF
CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU.
This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.

Table of Contents

Other manuals for DeVilbiss 7305 SERIES

Related product manuals