Disposable Collection Container with External Bacteria Filter
1. Remove bacteria filter, 3-3/8” tubing, and connection elbow and set aside.
2. Wash jar and lid in a solution of warm water with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water.
3. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131˚F-149ºF/55˚C-65˚C) solution for 60 minutes. Rinse with clean, warm water and air dry.
NOTE– The disposable collection container and lid are meant for single-patient use.
NOTE– The disassembled container may also be washed in a dishwasher, top shelf only, using a cycle with a water temperature between 131ºF-149ºF/55ºC-65ºC.
Reusable Collection Container with External Bacteria Filter
1. Remove bacteria filter, 4-3/8” tubing, and connection elbow and set aside. Remove o-ring and overflow valve from lid.
2. Wash jar, lid, o-ring, and overflow valve in a solution of warm water with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water.
3. For single patient use: After washing, disinfect using one of the following methods:
a. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) solution for 60 minutes. Rinse with clean, warm water and air dry in a clean
environment.
b. Wash with rubbing alcohol and air dry in a clean environment.
c. Wash with a commercial (bacterial-germicidal) disinfectant. Follow disinfectant manufacturer’s recommended dilution rates and instructions carefully.
4. For multi-patient use: After washing:
a. After parts are completely dry, place jar and lid in autoclave with the open end down. Ensure parts are not touching. Run one sterilization steam cycle at 250°F
(121°C) for 15 minutes. NOTE-Jar is guaranteed up to 30 cycles of autoclave sterilization at the indicated conditions.
b. Dispose of and replace filter, tubing and elbow between patients.
Suction Unit (single-patient use)
1. With the power switch in the “Off” position, disconnect the DeVilbiss Suction Unit from all external power sources.
2. Wipe the housing with a clean cloth and any commercial (bacterial-germicidal) disinfectant.
CAUTION– Do not submerge in water as this will result in damage to the vacuum pump.
NOTE–Do not use any cleaners or disinfectants that contain ammonia, benzene and/or acetone to clean the unit.
Tubing (single-patient use)
1. Disconnect the tubing from the unit.
2. Rinse thoroughly by running warm tap water through it.
3. Follow by soaking in a solution of 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) for 60 minutes. Rinse with clean, warm water and air dry.
4. Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.
Carrying Case (single-patient use)
1. Wipe the case using a clean cloth dampened with detergent and/or disinfectant.
Note– Disinfection information is based on AARC Clinical Practice Guideline Suctioning of the Patient in the Home.
Suction Unit (multi-patient use)
Device Cleaning and Disinfection When There is a Patient Change
When medical devices have already been used with a patient, contamination with human pathogenic germs should be assumed (unless there is evidence to the contrary), and the next
patient, user or third party should be protected by appropriate handling and preparation. Therefore, when there is a patient change, people must be protected during the transport and
handling of the device and the device must be fully processed, i.e., cleaned and disinfected, by suitably trained personnel before reuse to protect the next patient. The complete processing
may only be done by the manufacturer or by a qualified DeVilbiss provider/service technician.
Note – When the unit is used as per instructions under normal conditions the interior of the unit is protected from exposure to pathogens by the in-line filter on the collection container, hence
no disinfection of internal components is necessary.
Note – If the unit is used without an in-line filter then the interior of the unit has been exposed to pathogens and the unit cannot be disinfected.
Note – If the following processing of the unit by a qualified DeVilbiss provider/technician is not possible, the unit must not be used by another patient!
DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualified third party between uses by different patients.
1. Dispose of all accessory components that are not suitable for reuse, i.e., collection container, filter, tubing and carrying case.
2. With the power switch in the “Off” position, disconnect the DeVilbiss Suction Unit from all external power sources.
3. Visually inspect unit for any damage, missing parts, etc.
4. Wipe the housing with a clean cloth and a commercial (bacterial-germicidal) disinfectant that meets the requirements listed in the NOTE below and is used as per the disinfectant
manufacturer’s recommended dilution rates and instructions.
CAUTION – Do not submerge in water as this will result in damage to the vacuum pump.
NOTE – Do not use any cleaners or disinfectants that contain ammonia, benzene and/or acetone to clean the unit
Maintenance
Inspect suction tubing and container for leaks, cracks, etc. before each use.
DANGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is required, return unit to a qualified DeVilbiss Healthcare
provider or an authorized service center. Opening or tampering with the unit will void warranty.
Changing Filter Cartridge (single-patient use) Disposable Container