Guidance and Manufacturer’s Declaration – Emissions All Equipment and Systems
This device is intended for use in the electromagnetic environment specified below. The customer or user of this device should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Enforcement – Guidance
RF Emissions CISPR 11 Group 1
This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF Emissions CISPR 11
Class B Radiated and
Conducted Emissions
This device is suitable for use in all establishments including domestic, and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonics IEC 61000-3-2 Class A
Flicker IEC 61000-3-3 Complies
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment – Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or ceramic tile. If floors are synthetic, the relative
humidity should be at least 30%.
Electrical Fast Transient/burst
IEC 61000-4-4
±2kV on AC Mains ±2kV on AC Mains Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC 61000-4-
11
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
Mains power quality should be that of a typical commercial or hospital environment. If
the user of this device requires continued operation during power mains interruptions, it
is recommended that the device be powered from an uninterruptible power supply or
battery.