EN - 11SE-DV54-HE
ACCESSORIES/REPLACEMENT ITEMS
Replacement Items
Air-inletlter(4/pk) DV51D-602 Heater connection cover
DV51D-605
Fineparticlelter(4/pk) DV51D-603 Oxygen adapter
7353D-601
Air supply tubing (6’) CPAPTUB6-90 Carrying case
DV51D-610
Air supply plug DV51D-604 HeatedHumidier
DV5HH
HumidierChamber DV5C Chamber Sealing Gasket
DV5C-614
Power Items
AC power cord (USA) DV51D-606 AC power cord (Australia) DV51D-609
AC power cord (UK) DV51D-608 DC power cord DV51D-619
AC power cord (EU) DV51D-607
DC battery clamp-on adapter (requires DC
Power Cord DV51D-619)
DV51D-696
MAINTENANCE
Expected Service Life:
• CPAPandhumidierbase=3years
• Humidierwaterchamber=1Year
• Tubingandgrossparticleintakelter=180days
DANGER
Electric shock hazard – Do not attempt to open or remove the cabinet. There are no user-serviceable components
inside. If service is required, contact your equipment provider for instructions on obtaining service. Opening or
attempting to service your device will void the warranty.
WARNING
Disconnect supply before servicing.
Pressure Accuracy–The SleepCube is designed to provide accurate pressure performance from the factory without
additional calibration. No routine service is required provided the device is used in accordance with the manufacturer’s
directions.
NOTE–Some countries require periodic servicing and calibration of this kind of medical device. Contact your equipment
provider for further information.
Standard Air-Inlet Filter–Checkthelterevery10daysandcleanitifnecessary.Replacethelterevery6monthsor
soonerifitisdamaged.Thestandardlterisdesignedtolterparticleslargerthan3.0micronsinsize.TheSleepCube
MUSThavethestandardlterinplaceduringoperation.
Optional Fine Particle Filter–Checkthelterevery10daysandreplaceitifitisdirtyordamaged,otherwisereplacethe
lterevery30days.Theoptionalneparticlelterisdesignedtolterparticlesassmallas0.3micronsinsize.
Maintenance:
Inspection Interval Replacement Interval Specication
Gross Particle Intake Filter 10 days, clean if dirty 6 months > 3.0 micron particles
Fine Particle Intake Filter 10 days, replace if dirty 30 days > 0.3 micron particles
Patient tubing Daily 6 months See tubing spec
Humidierchamber 1 / month 1 year