EN - 4 SE-DV54-HE
SYMBOL DEFINITIONS
Class II electrical protection-
double insulated
Locked Standby-turns blower on/off
Data port input/output
Unlocked Heat
Next item on LCD display IPX1 Drip-proof, vertical 100-240 VAC input 50/60 Hz
Previous item on LCD display
DC input: 12 VDC, center pin
positive
Delay-starts the delay cycle
Increase value on LCD
display
Attention, consult
accompanying documents
Manufacturer
Decrease value on LCD
display
Type BF equipment-applied
part
The device contains electrical and/or electronic equipment that must be recycled per EU Directive 2012/19/EU -
Waste Electrical and Electronic Equipment (WEEE)
IMPORTANT SAFEGUARDS
READ ALL INSTRUCTIONS BEFORE USING
THIS DEVICE.
SAVE THESE INSTRUCTIONS.
DANGER
Electric Shock Hazard – Do not use while bathing.
Electric Shock Hazard – Do not immerse this device into water or any other liquid.
Electric Shock Hazard – Do not attempt to open or remove the cabinet. There are no user-serviceable internal
components. If service is required, return the product to your home care provider. Opening or tampering with the
product will void the warranty.
WARNINGS
• WARNING-APPLIEDPARTSTHATMAYEXCEED41˚C.
Thefollowingsurfacetemperaturesmayexceed41˚Cundercertainconditions:
External surface of the CPAP .............................. 42.7˚C
Patient Tubing Connector .................................... 43.9˚C
LCD Display ......................................................... 42.0˚C
On/Off & Delay Button ......................................... 41.8˚C
• HumidierHeaterPlate ....................................... 60.6˚C
Serial Communication Port .................................. 42.7˚C
The DeVilbiss SleepCube should be used only with masks recommended by DeVilbiss, your physician or respiratory
therapist.
To avoid rebreathing of exhaled air, do not use a CPAP mask unless the device is turned on and providing a supply
of air. Venting in the mask should never be blocked. When the device is turned on and providing a fresh supply of
air,exhaledairisushedoutofthemaskvent.However,whenthedeviceisnotoperating,exhaledairmaybe
rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some circumstances, lead to
suffocation. This warning applies to most CPAP devices.
The DeVilbiss SleepCube is not a life support device and may stop operating with certain device faults or a power
failure. It is intended to be used on spontaneously breathing individuals weighing 66 lbs/30 Kg or greater.
To avoid electric shock, always unplug power cord from wall outlet power source when performing cleaning.
Use only accessories recommended by DeVilbiss.
Alwaysremovethewaterchamberfromtheoptionalhumidiercradlebeforelling.
Ifwaterhasspilledontothehumidiercradleorithasbeensubmergedintowater,unplugpowercordfrompower
source immediately. Allow device to dry completely before use.
Neverblockairopeningsofthehumidiercradleorchamber.Donotinsertobjectsintoanyopeningsortubes.
Forproperoperation,placesystemonaat,sturdy,levelsurface.
Theoptionalhumidierisintendedforsingle-patientuseonly.