Humidifier
Humidity Output (in the operating flow range): ................................................................................................ ≥10mgH
2
O/l air
Size: ......................................................................................................2.6” H x 6.3” W x 8.4” D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: ........................................................................................................................................................1.75 lbs. (0.794 Kg)
Power Consumption: .................................................................................................................................................... 85 Watts
DV5HH Electrical Requirements ................................................................................................................100-240VAC, 0.95 A
Heater Plate Thermal Cutout:..............................................................................................................................305°F (152°C)
Power/Temperature Control: ......................................................................................Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: ...............................................................................84°F to 149°F (approx. 29°C to approx. 65°C)
Operating Temperature Range: .................................................................................................... 41°F to 104°F (5°C to 40°C)
Operating Humidity Range: ....................................................................................................... 0 to 95% R.H. non-condensing
Operating Atmospheric Pressure Range: ........................................................................................................70.0 – 106.0 kPA
Filter Specifications
Standard Filter ......................................................................................................................................... > 3.0 micron particles
Optional Fine Particle Filter ..................................................................................................................... > 0.3 micron particles
Dynamic Short-Term Pressure Accuracy with and without DV5HH Humidifier (Per
ISO 17510-1:2007)
Waveform Volume [mL] Breath Rate [min-1]
Set Pressure
[cmH
2
O]
Pressure Accuracy
[Pk-Pk cmH
2
O]
Sinusoidal
cyclewithI:E=
1:1
500 10, 15, 20
3.0 0.5
7,0 0.5
12,0 1.0
16,0 1.0
20,0 1.0
Maximum Flow-Rate (per ISO 17510-1:2007)
Test Pressures
3,0 cmH
2
O 7,0 cmH
2
O 12,0 cmH
2
O 16,0 cmH
2
O 20,0 cmH
2
O
Measured pressure at the patient
connection port (cmH
2
O)
2.9 6.4 11.0 15.0 19.0
Averageowatthepatientconnectionport
(l/min)
83.8 126.8 137.4 130.0 113.8
Long Term Static Pressure Accuracy ................................................................................................................. +/- 0.5 cmH
2
O
Additional Specifications
Equipment classification with respect to protection from electric shock ........................................................................ Class II
Degree of protection from electric shock ................................................................................................. Type BF Applied Part
Degree of protection against ingress of liquids ...................................................................................... IPXI Drip-proof vertical
Mode of operation .................................................................................................................................................... Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.

Related product manuals