EN - 10 SE-DV54-1
Supplemental oxygen is sometimes prescribed in addition to CPAP for treatment of Obstructive Sleep Apnea. If your
physician prescribed oxygen for your sleep therapy, you can add the oxygen in one of two ways:
1. Obtain the optional oxygen adapter (part #7353D-601) from your homecare provider and connect it to the air supply
port on the SleepCube. Connect one end of the air supply tubing to the open end of the adapter and the other end of
the air supply tubing to the mask.
2. If your mask is equipped with an oxygen port, your physician may request that you apply oxygen directly to your mask.
ACCESSORIES/REPLACEMENT ITEMS
Replacement Items
Air-inletlter(4/pk) DV51D-602 Heater connection cover DV51D-605
Fineparticlelter(4/pk) DV51D-603 Oxygen adapter 7353D-601
Air supply tubing (6’) DV51D-629 Carrying case DV51D-610
Air supply plug DV51D-604 HeatedHumidier DV5HH
HumidierChamber DV5C Chamber Sealing Gasket DV5C-614
Power Items
AC power cord (USA) DV51D-606 AC power cord (Australia) DV51D-609
AC power cord (UK) DV51D-608 DC power cord DV51D-619
AC power cord (EU) DV51D-607
DC battery clamp-on adapter (requires DC
Power Cord DV51D-619)
DV51D-696
DeVilbiss Masks
FlexSet
®
Masks Serenity
®
Masks
Large, Silicone 9354L Large, Silicone 9352L
Medium, Silicone 9354D Medium, Silicone 9352D
Small, Silicone 9354S Small, Silicone 9352S
Large, Gel 9354GL Large, Gel 9352GL
Medium, Gel 9354G Medium, Gel 9352G
Small, Gel 9354GS Small, Gel 9352GS
Large, ComfortTouch
™
Silicone 9354LR Large, ComfortTouch Silicone 9352LR
Medium, ComfortTouch Silicone 9354DR Medium, ComfortTouch Silicone 9352DR
Small, ComfortTouch Silicone 9354SR Small, ComfortTouch Silicone 9352SR
Large, ComfortTouch Gel 9354GLR Large, ComfortTouch Gel 9352GLR
Medium, ComfortTouch Gel 9354GR Medium, ComfortTouch Gel 9352GR
Small, ComfortTouch Gel 9354GSR Small, ComfortTouch Gel 9352GSR
MAINTENANCE
DANGER
Electricshockhazard-Donotattempttoopenorremovethecabinet;therearenouser-serviceablecomponents
inside.Ifserviceisrequired,contactyourequipmentproviderforinstructionsonobtainingservice.Openingor
attemptingtoserviceyourdevicewillvoidthewarranty.
WARNING
Disconnectsupplybeforeservicing.
PressureAccuracy–The SleepCube is designed to provide accurate pressure performance from the factory without
additional calibration. No routine service is required provided the device is used in accordance with the manufacturer’s
directions.
NOTE–Some countries require periodic servicing and calibration of this kind of medical device. Contact your equipment
provider for further information.
StandardAir-InletFilter–Checkthelterevery10daysandcleanitifnecessary.Replacethelterevery6monthsor
soonerifitisdamaged.Thestandardlterisdesignedtolterparticleslargerthan3.0micronsinsize.TheSleepCube
MUSThavethestandardlterinplaceduringoperation.