EN - 11SE-DV54-1
OptionalFineParticleFilter–Checkthelterevery10daysandreplaceitifitisdirtyordamaged;otherwisereplacethe
lterevery30days.Theoptionalneparticlelterisdesignedtolterparticlesassmallas0.3micronsinsize.
CLEANING
WARNING
Toavoidelectricshock,alwaysunplugthepowercordfromthewalloutletpowersourcebeforecleaning.
CAUTION–Never rinse or place the device in water. Never allow liquids to get into or around any of the ports, switches or
air lter; doing so will result in product damage. If this occurs, do not use the device. Contact your equipment provider for
service.
Device Cabinet
1. Unplug the SleepCube from the power source. Wipe the cabinet with a clean, damp cloth every few days to keep the
cabinet dust free.
2. Allow the device to dry completely before plugging it into a power source.
Filters
Air-InletFilter
1. Theair-inletltershouldbecheckedevery10daysandcleanedasnecessary.Thelterislocatedintherearofthe
device.
CAUTION–Proper lter function is important for the operation of the device and to protect the device from damage.
2. Toclean,removethedarkouterfoamlterfromtheframe.
3. Washthelterinasolutionofwarmwateranddishwashingdetergent,andrinsewithwater.Allowtheltertodry
completely.Ifthelter(part#DV51D-602)isdamaged,contactyourequipmentproviderforareplacement.
4. Ensurethelteriscompletelydrybeforeinstallationanduse.
FineParticleFilter(Optional)
1. Ifyouhaveanoptionalneparticlelter(part#DV51D-603),inspectthislterandreplaceitifitshowssignsofdirtor
damage. Do not wash.
2. Installtheneparticlelterbehindthestandardlter.
NOTE–If the ne particle lter is not installed rst according to the directions, the life of the lter will be reduced requiring
more frequent replacement.
Optional Heated Humidifier
WaterChamber-Humidierwaterchambershouldbecleaneddaily.
1. Press on the chamber release latch and pull out the water chamber from the cradle (Fig 4).
2. Separate the chamber halves by pressing forward on the chamber disassembly lever and lifting the lid of the
chamber (Fig 9).
3. Remove the sealing gasket from the chamber lid (Fig 10).
4. Hand wash the parts in a solution of warm water and mild dishwashing detergent.
CAUTION–Do not use any solutions that contain bleach, chlorine, alcohol, scented oils or other aromatic-based ingredients
to clean the humidier components or tubing. These solutions may reduce the life of the product.
5. Allow the parts to dry before reassembly.
CAUTION–Inspect the gasket seal before each use to ensure that it is properly seated in the groove around the entire
chamber.
NOTE– The disassembled chamber may also be washed once per week in a dishwasher on the delicate or glassware
cycle, top shelf only.
HumidierCradle-Thehumidiercradleshouldbecleanedasneeded.
1. Turn off the heater and unplug from power source. Allow system to cool for at least 10 minutes.
2. Detachowgeneratorfromthecradle;rstremovewaterchamberthenfullydepresstheowgeneratorrelease
button to separate (Fig 4).
NOTE–Flow generator release button must be pressed completely to separate.
3. Use a soft, damp cloth to clean the surfaces of unit.