EN - 13SE-DV54-1
Issue Possible Cause Remedy
Watercondensation
collectinginthe
hosecausinga
gurglingsound.
1.Humidicationistoohigh. 1.Reducehumidierheatersetting.
2.Roomtemperatureuctuatesfromhigherto
lower levels through the night.
2. Increase room temperature.
Airfromow
generatorseemsto
betoowarm.
1.Airltersaredirty. 1.Cleanlters.
2. Air inlet port is blocked. 2. Unblock air inlet.
3. Room temperature is too high. 3. Lower room temperature.
4. Device is located near a heat source. 4. Locate device away from heat source.
5.Humidier(optional)heatersettingistoo
high.
5. Lower heater setting.
Nasal,sinusorear
pain,runnynose
Youmayhaveareactiontoairowpressure. Discontinue use and contact physician.
Heatersettingis
notvisibleon
display
1.Theowgeneratorisnotfullyseatedonto
thehumidiercradle.
1. Ensure proper contact is made (Fig. 3).
2. Your CPAP device is not equipped with the
optionalhumidier.
SPECIFICATIONS
CPAP
Size ........................................................................................................... 4.2”H x 6.5”W x 6.9”D (10.7cm x 16.5cm x 17.5cm)
Weight ............................................................................................................................................................. 1.22 kg (2.7 lbs.)
Electrical Requirements AC ...................................................................................................................... 100-240V~, 50/60 Hz
Electrical Requirements DC ..................................................................................................................10.4 - 15 VDC, 5 Amps
Maximum Power Consumption ......................................................65 watts max from AC power source (flow generator only)
Pressure Range ......................................................................................................................................................3-20 cmH
2
O
Operating Temperature Range ....................................................................................................... 41˚Fto104˚F(5˚Cto40˚C)
Operating Humidity Range .......................................................................................................0%to95%RHnon-condensing
Operating Atmospheric Conditions ..........................................................................................Sea level to 9,000 feet (2743 m)
Storage & Transportation Temperature Range ..................................................................... -40°F to +158°F (-40°C to 70°C)
Storage & Transportation Humidity Range ...............................................................................0%to95%RHnon-condensing
Maximum Limited Pressure ...........................................................................................................20 cmH
2
O under normal use
Sound Pressure Level (tested per ISO 17510) ...............................................................................................................26 dBA
Sound Power Level .........................................................................................................................................................38 dBA
Humidifier
Humidity Output (in the operating flow range): ................................................................................................ ≥10mgH
2
O/l air
Size: ......................................................................................................2.6” H x 6.3” W x 8.4” D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: ........................................................................................................................................................1.75 lbs. (0.794 Kg)
Power Consumption: .................................................................................................................................................... 85 Watts
DV5HH Electrical Requirements ................................................................................................................100-240VAC, 0.95 A
Heater Plate Thermal Cutout:..............................................................................................................................305°F (152°C)
Power/Temperature Control: ......................................................................................Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: ...............................................................................84°F to 149°F (approx. 29°C to approx. 65°C)
Operating Temperature Range: ....................................................................................................41°F to 104°F (5°C to 40°C)
Operating Humidity Range: ....................................................................................................... 0to95%R.H.non-condensing
Operating Atmospheric Pressure Range: ........................................................................................................70.0 – 106.0 kPA