Issue Possible Cause Remedy
Watercondensation
collectinginthe
hosecausinga
gurglingsound.
1.Humidicationistoohigh. 1.Reducehumidierheatersetting.
2.Roomtemperatureuctuatesfromhigherto
lower levels through the night.
2. Increase room temperature.
Airfromow
generatorseemsto
betoowarm.
1.Airltersaredirty. 1.Cleanlters.
2. Air inlet port is blocked. 2. Unblock air inlet.
3. Room temperature is too high. 3. Lower room temperature.
4. Device is located near a heat source. 4. Locate device away from heat source.
5.Humidier(optional)heatersettingistoo
high.
5. Lower heater setting.
Nasal,sinusorear
pain,runnynose
Youmayhaveareactiontoairowpressure. Discontinue use and contact physician.
Heatersettingis
notvisibleon
display
1.Theowgeneratorisnotfullyseatedonto
thehumidiercradle.
1. Ensure proper contact is made (Fig. 3).
2. Your CPAP device is not equipped with the
optionalhumidier.
SPECIFICATIONS
CPAP
Size ........................................................................................................... 4.2”H x 6.5”W x 6.9”D (10.7cm x 16.5cm x 17.5cm)
Weight ............................................................................................................................................................. 1.22 kg (2.7 lbs.)
Electrical Requirements AC ...................................................................................................................... 100-240V~, 50/60 Hz
Electrical Requirements DC ..................................................................................................................10.4 - 15 VDC, 5 Amps
Maximum Power Consumption ......................................................65 watts max from AC power source (flow generator only)
Pressure Range ......................................................................................................................................................3-20 cmH
2
O
Operating Temperature Range ....................................................................................................... 41˚Fto104˚F(5˚Cto40˚C)
Operating Humidity Range .......................................................................................................0%to95%RHnon-condensing
Operating Atmospheric Conditions ..........................................................................................Sea level to 9,000 feet (2743 m)
Storage & Transportation Temperature Range ..................................................................... -40°F to +158°F (-40°C to 70°C)
Storage & Transportation Humidity Range ...............................................................................0%to95%RHnon-condensing
Maximum Limited Pressure ...........................................................................................................20 cmH
2
O under normal use
Sound Pressure Level (tested per ISO 17510) ...............................................................................................................26 dBA
Sound Power Level .........................................................................................................................................................38 dBA
Humidifier
Humidity Output (in the operating flow range): ................................................................................................ ≥10mgH
2
O/l air
Size: ......................................................................................................2.6” H x 6.3” W x 8.4” D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: ........................................................................................................................................................1.75 lbs. (0.794 Kg)
Power Consumption: .................................................................................................................................................... 85 Watts
DV5HH Electrical Requirements ................................................................................................................100-240VAC, 0.95 A
Heater Plate Thermal Cutout:..............................................................................................................................305°F (152°C)
Power/Temperature Control: ......................................................................................Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: ...............................................................................84°F to 149°F (approx. 29°C to approx. 65°C)
Operating Temperature Range: ....................................................................................................41°F to 104°F (5°C to 40°C)
Operating Humidity Range: ....................................................................................................... 0to95%R.H.non-condensing
Operating Atmospheric Pressure Range: ........................................................................................................70.0 – 106.0 kPA

Table of Contents

Other manuals for DeVilbiss DV5HH

Related product manuals