EN - 9SE-DV54-1
MaskLeak–ThismessagemeanstheSleepCubehasdetectedalargeamountofairowduringtherapyforatleast10%
ofthetimeduringyourprevioususesession.Thisnoticationisdisplayedwhenthedeviceisturnedon.Ifthismessageis
displayed,putonthemaskandadjusttheheadgeartoensurethemaskisproperlyttedtoyourface.Followthemask
manufacturer’sdirectionsforadjustingthemaskandheadgearforpropertting.
MaskOff–Thismessageappearsduetoapoormasktoraremovedmask.Checkforleakaroundthemasksealand
makeadjustmentsasnecessaryaccordingtothemaskmanufacturer’sinstructions.Thismessagewillbedisplayeduntil
thehighairowproblemiscorrected.Ifthehighairowconditionpersistsforabout20seconds,thedevicewill
automatically turn off if Auto-OFF is enabled.
DeviceFault – If a device fault message is displayed, refer to Troubleshooting for instructions.
SLEEPCUBE TRAVEL INFORMATION
International Power Changes
Your DeVilbiss SleepCube is equipped with a universal power supply. This means that it is automatically capable of
acceptinglinevoltagesof100to240V~50/60Hz.Youdonotneedtomakeanyadjustmentstothedeviceifyouare
traveling to another country where the line voltage is different than your domestic line voltage. Simply contact your
equipment provider for the correct style power cord for the region in which you will be traveling (refer to Accessories/
Replacement Items for the correct part number).
Altitude
Your SleepCube automatically compensates for changes in altitude between sea level and 9000 ft (2750 m). You do not
needtomakeadjustmentstothedeviceforchangesinaltitude.
SLEEPCUBE DC OPERATION
Battery Power
Your SleepCube can be operated from a 12V DC power source in a motor home, boat or other recreational vehicle using
the 12V DC cable. Alternatively you can power your SleepCube using a 12V DC stand alone battery with a DC battery
clamp on adapter and 12V DC cable. All cables are available from your equipment provider (refer to Accessories/
Replacement Items for the correct part numbers). The stand alone battery connection uses clamps that attach to the
terminal connections of the battery.
NOTE–The heater for the SleepCube’s humidication system will not operate directly from a 12V DC power source. If you
need humidication, non-heated humidication is still available. Alternately, instead of connecting12V DC power directly to
the SleepCube, you can pass the DC power through an inverter that will deliver AC power to the system. This will allow the
heater to function. The power rating of the inverter must be at least 200 watts @ 110V~ / 400 watts @ 220V~.
NOTE–The SleepCube will automatically switch power sources between AC and 12V DC if both sources are available. If
your unit is connected to both AC and 12V DC power sources, it will operate on AC power as long as AC power is present,
and automatically switch to 12V DC power if AC power is lost. When AC power is restored, it will return to the AC source.
You do not need to reset or adjust any settings when the power source switches, but remember that your heater will not
operate on 12V DC.
SUPPLEMENTAL OXYGEN
WARNING-OXYGEN USE
• Oxygensupportscombustion.Toavoidpossiblephysicalinjury,donotsmokewhileusingthisdevicewith
supplementaloxygen.Donotusethisdevicenearhotobjects,volatilesubstancesorsourcesofopenames.
• Always turn on the SleepCube before turning on the oxygen source. Turn off the oxygen source before turning off the
SleepCube.
• Thisdeviceisnotsuitableforuseinthepresenceofaammableanestheticmixture.
• Ataxedowrateofsupplementaloxygen,theinhaledoxygenconcentrationwillvarydependingonthepressure
setting, patient breathing pattern, mask selection, and leak rate. This warning applies to most types of CPAP devices.
• Never allow the oxygen source to run continuously while connected to the SleepCube if the device is not in use. If
theSleepCubeisnotbeingused,turnofftheoxygenow.
CAUTION
• Oxygen is prescribed gas and should only be administered under the supervision of a physician.
• Theowsettingoftheoxygensourcemustbespeciedbyaphysician.